Products

Model:
HD-CLZ1
Introduction:
Online Airborne Particle Counter measures dust particles in six size channels online and judges the cleanliness level per GB/T6167-2007. Widely used for cleanroom, workshop and pharmaceutical clean-area online monitoring....
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
The intelligent airborne particle counter is an instrument whose confidence-level measurement and verification standard follows the conditions specified in the national standard GB/T 6167-2007 and the calibration specification JJF 1190-2008, and which automatically determines the cleanliness level. One sampling can simultaneously measure the number of dust particles in six particle size channels and simultaneously display the number and variation of particles in the six channels.
It satisfies the cleanliness testing needs of the pharmaceutical industry (pharmaceutical factories, drug inspection institutes, hospitals, etc.); the electronics industry (semiconductor factories, precision machinery production and processing, etc.); the food hygiene industry (fine processing of dairy products, plastic-sealed meat products, condiments, agricultural products, etc.); manufacturers of filters (inspecting filter quality and efficiency, etc.); and cleanrooms (zones) required for precision processing, precision testing and bioengineering in optics, aerospace and other fields.
| Parameter | Specification |
|---|---|
| Sampling mode | Timed detection, continuous detection |
| Sampling volume | 2.83 L/min (0.1 cfm) |
| Six particle size channels | 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 (μm) |
| Maximum sampling concentration | 35000 particles/L (particle size ≥ 0.5 μm) |
| Self-cleaning time | ≤ 10 min |
| Particle size distribution error | 0.5 μm: ≤ ± 30% |
| Particle concentration indication error | 0.5 μm: ≤ ± 30% |
| Measurement repeatability relative error | < 10% |
| Flow error | ≤ ± 5% |
| Cleanliness standards | National standard, ISO, 209E, GMP |
| Operating environment | Working environment 0-40℃, 10-75%RH |
| Sampling period | 10 s to 99 h |
| Sampling delay | ≤ 60 s |
| Measurement interval | ≤ 60 s |
| Data communication | Standard USB, Wi-Fi communication, RS485, LAN |
| Light source | Imported semiconductor laser diode (service life more than 30,000 hours) |
| Humidity | Range 0-100%RH, error ± 2%RH, repeatability 0.1%RH |
| Temperature | Range -40 to 85℃, error ± 0.3℃, repeatability 0.1℃ |
| Atmospheric pressure | Range 300-1100 hPa, accuracy ± 1 hPa, resolution 0.1 hPa |
| Display | 7-inch color touch screen |
| Data memory capacity | 10,000-100,000 records, queryable and exportable |
| Power parameters | 12V / 3 A |
| Power | 30 W |
| Main unit dimensions | 210 × 150 × 65 mm (L × W × H) |
| Main unit weight | 1.8 kg |
| Packaging dimensions | 350 × 150 × 300 mm |
| Packaging weight | 2.8 kg |