COD Digester performs reflux digestion of water samples for COD determination per HJ828-2017. Widely used for environmental monitoring, soil analysis, food safety testing and disease control....
The intelligent COD reflux digester follows the basic principle of the national standard HJ 828-2017 "Water quality - Determination of chemical oxygen demand - Dichromate method". It guarantees the two-hour reflux digestion at micro-boiling temperature, and the preparation and addition of the reagent solutions are identical to those of the standard method, which ensures reliable and accurate analytical results.
It is widely used for the pretreatment of water samples in environmental monitoring, soil analysis, food safety testing and centres for disease control and prevention.
| 1 | Efficient and accurate | standardized operating procedures and precisely controlled test conditions deliver fast digestion with consistent and accurate results. |
| 2 | Simple operation and high automation | an automatic control system allows the temperature and time to be set with one key, which reduces human error and simplifies the workflow. |
| 3 | Excellent safety | a glass serrated reflux tube replaces the spherical reflux tube, and air cooling combined with water cooling replaces the traditional tap-water cooling, saving water and standardizing the procedure; over-temperature power cut-off further improves safety. |
| 4 | Easy-to-clean panel | the graphene heating plate has a smooth, flat surface, so stains can be wiped off immediately and damage caused by careless operation is reduced. |
| 5 | Environmentally friendly | microcomputer timing control heats the whole conical-flask reflux assembly, saving energy, reducing the electrical load and improving efficiency. |
| 6 | Visual status feedback | status indicator lamps on the main unit show the operating state by colour, which is intuitive and convenient. |
| Heating method | Far-infrared black crystal panel |
| Batch capacity | 6 samples |
| Heating flask | 250 mL, 24/29 ground-joint flat-bottom conical flask with serrated reflux tube |
| Temperature range | Room temperature to 350 °C |
| Measurement method | HJ 828-2017 "Water quality - Determination of chemical oxygen demand - Dichromate method" |
| Timing range | 0–999 min |
| Measurement error | Potassium hydrogen phthalate standard solution (500 mg/L): relative standard deviation ≤ 5.0%; industrial wastewater (500 mg/L): relative standard deviation ≤ 8.0% |
| Cooling method | Air cooling + water cooling |
| Status feedback | Indicator lamp colour shows the operating state |
| Ambient temperature | 5–40 °C |
| Ambient humidity | ≤85% RH (non-condensing) |
| Power | 2000 W |
| Power supply | AC 220 V / 50 Hz |
| Dimensions | 448 × 406 × 565 mm |
| Weight | 12 kg |