EN
Position:Home>Products>Sample Pretreatment>Nitrogen Evaporator>Automatic>Automatic Nitrogen Evaporator

Products

Automatic Nitrogen Evaporator

Model:

HD-QD24S

Introduction:

Automatic Nitrogen Evaporator (Nitrogen Concentrator) concentrates samples quickly by nitrogen blowing for environmental and food safety analysis. Widely used for pesticide residue, pharmaceutical and biological sample pretreatment....

  • Image and Text Display
  • Technical Specifications
  • Product video

I. Product Description

In a large amount of analytical work, especially in the fields of environmental pollutants and food-safety analysis, samples must be pre-treated to obtain trace target components. The process mainly includes sample extraction, concentration, purification and re-concentration, among which rapid and lossless concentration is a key step. Common concentration devices include rotary evaporators, K-D concentrators and nitrogen purge (nitrogen concentrator) units; nitrogen purge concentration is the simplest. When a large batch of samples must be concentrated simultaneously and the solvent volume is large, the above devices become inadequate — the process is time-consuming and laborious (requires supervision to prevent drying), and the open nitrogen purge device is harmful to operators' health.

This instrument uses a unique concentration method to greatly increase the concentration rate. Meanwhile, it uses a built-in exhaust fan to discharge the evaporated waste gas through an exhaust pipeline, so that a nitrogen purge concentration device that would otherwise have to be placed in a fume hood can be safely installed on an ordinary laboratory bench. It is easy to move, saves laboratory cost, and reduces the harm of toxic and harmful solvents to operators. It is a laboratory sample pre-treatment device.

II. Product Characteristics

The automatic nitrogen concentrator can concentrate single or multiple samples simultaneously without manual attendance: it uses multiple optical sensors to monitor the concentration process of each sample; when the sample is evaporated to the preset volume, the system automatically stops the nitrogen purging of the corresponding channel and gives an alarm. The whole concentration process needs no manual supervision.

Special airflow purging trajectory and buffering design: accelerates solvent evaporation and concentration and prevents splashing loss of solvent.

Working parameters can be set arbitrarily, controlled and displayed in real time: the main working parameters — nitrogen purge pressure, water bath temperature and working time — can all be set as required.

Stable and constant nitrogen purge airflow pressure: the instrument has a built-in automatic pressure-regulating device; the airflow pressure can be automatically controlled and kept constant, unaffected by the sudden opening, closing or number of working channels (samples).

No sample contamination: all gas paths and related components use validated zero-contamination materials, avoiding contamination of the sample by the instrument.

Simple and safe operation: flexible working-parameter setting and convenient sample loading/removal; easy to learn and use. The fully enclosed design and the instrument's powerful exhaust system avoid the influence of water-bath steam and organic volatiles on the instrument and operators.

III. Technical Parameters

ItemSpecification
Product ModelHD-QD24S
Sample Quantity1-24 samples concentrated simultaneously
Sample Vial Volume50 or 150 mL
Endpoint DetectionEach working channel is equipped with an optical sensor for automatic, independent endpoint detection
Endpoint Volume1.0 mL, 0.5 mL or near-dry (~0.1 mL; with appropriately extended purging time the solvent can also be blown dry); concentrating vials of different specifications can be used crosswise at the same time
Water Bath TemperatureRoom temperature to 95℃ (±0.5℃)
Nitrogen Purge Time0-9999 s
Gas PressurePurge working pressure 0~0.1 MPa (pressure step 0.01 MPa)
External Nitrogen Pressure Range0.2~0.8 MPa; external allowable pressure 1.0 MPa
Gas ConsumptionApprox. 500 mL/min per channel at purge pressure (0.1 MPa) (approx. 17 cfm)
Constant-Volume Sensitivity10 levels adjustable; accurate concentration and constant-volume for solvents of different colors or transmittance
Control ModeManual or intelligent control of the purge endpoint as required
AlarmAutomatic alarm on lid opening, completion of concentration, or insufficient water-bath level or nitrogen pressure
Power Supply220V/60Hz
Instrument Dimensions740×460×345 mm
Weight20 kg
The video is currently in production, please stay tuned......

Shandong Horde Electronic Technology Co., Ltd.

  • Phone:008619053605658

  • Email:salesmanager@hordeinstrument.com

whatsapp

Copyright © Shandong Horde Electronic Technology Co., Ltd. All Rights Reserved.
Disclaimer: This website cannot identify the intellectual property rights of uploaded images or text. If there is any infringement, please notify us in a timely manner, and this website will review and delete it as soon as possible.

Phone

008619053605658

Contact
Top