Products

Model:
DSC-Q8
Introduction:
Differential Scanning Calorimeter measures glass transition temperature, phase transition, melting enthalpy, product stability and oxidation induction time per GB/T 19466 and ISO 11357. Widely used for polymer, pharmaceutical and material research....
The DSC-Q8 differential scanning calorimeter is a touchscreen-operated differential scanning calorimeter that can perform glass transition temperature testing, phase transition testing, melting and enthalpy value testing, product stability testing and oxidation induction time testing.
GB/T 19466.2-2004 / ISO 11357-2:1999 Part 2: Determination of glass transition temperature
GB/T 19466.3-2004 / ISO 11357-3:1999 Part 3: Determination of melting and crystallization temperature and enthalpy
GB/T 19466.6-2009 / ISO 11357-3:1999 Part 6: Determination of oxidation induction time and oxidation induction temperature
| Temperature range | Room temperature ~ 500 °C |
| Temperature resolution | 0.01 °C |
| Temperature fluctuation | ±0.1 °C |
| Temperature repeatability | ±0.1 °C |
| Heating rate | 0.1~100 °C/min |
| Constant temperature time | Recommended <24 h |
| Temperature control mode | Heating, constant temperature (fully automatic program control) |
| DSC range | 0 ~ ±600 mW |
| DSC resolution | 0.01 mW |
| DSC sensitivity | 0.01 mW |
| Power supply | AC 220 V / 50 Hz or customized |
| Atmosphere control gas | Nitrogen, oxygen (automatically switched by the instrument) |
| Gas flow rate | 0-300 mL/min |
| Gas pressure | 0.2 MPa |
| Display | 15-inch LCD touchscreen |
| Data interface | Standard USB interface; standard substances (indium, tin, zinc) included, allowing users to calibrate temperature themselves |
1. Industrial-grade 15-inch touchscreen, completely free from the constraints of a computer.
2. Electronic flow meter for more accurate flow recording.
3. Touch control interface independent of the display, more flexible.
4. Unique furnace sealing provides excellent experimental conditions.
5. Automatic switching of two atmosphere gas lines, fast switching speed and short stabilization time; an additional protective gas input is provided.
6. Simple and easy-to-operate software.
7. Standard substances (indium, tin, zinc) are provided for user self-calibration of temperature.