EN
Position:Home>Products>Physical Performance Testing>DSC>Differential Scanning Calorimeter

Products

Differential Scanning Calorimeter

Model:

DSC-Q8

Introduction:

Differential Scanning Calorimeter measures glass transition temperature, phase transition, melting enthalpy, product stability and oxidation induction time per GB/T 19466 and ISO 11357. Widely used for polymer, pharmaceutical and material research....

  • Image and Text Display
  • Technical Specifications
  • Product video

I. Product Introduction

The DSC-Q8 differential scanning calorimeter is a touchscreen-operated differential scanning calorimeter that can perform glass transition temperature testing, phase transition testing, melting and enthalpy value testing, product stability testing and oxidation induction time testing.

II. Applicable Standards

GB/T 19466.2-2004 / ISO 11357-2:1999 Part 2: Determination of glass transition temperature

GB/T 19466.3-2004 / ISO 11357-3:1999 Part 3: Determination of melting and crystallization temperature and enthalpy

GB/T 19466.6-2009 / ISO 11357-3:1999 Part 6: Determination of oxidation induction time and oxidation induction temperature

III. Technical Parameters

Temperature rangeRoom temperature ~ 500 °C
Temperature resolution0.01 °C
Temperature fluctuation±0.1 °C
Temperature repeatability±0.1 °C
Heating rate0.1~100 °C/min
Constant temperature timeRecommended <24 h
Temperature control modeHeating, constant temperature (fully automatic program control)
DSC range0 ~ ±600 mW
DSC resolution0.01 mW
DSC sensitivity0.01 mW
Power supplyAC 220 V / 50 Hz or customized
Atmosphere control gasNitrogen, oxygen (automatically switched by the instrument)
Gas flow rate0-300 mL/min
Gas pressure0.2 MPa
Display15-inch LCD touchscreen
Data interfaceStandard USB interface; standard substances (indium, tin, zinc) included, allowing users to calibrate temperature themselves

IV. Technical Features

1. Industrial-grade 15-inch touchscreen, completely free from the constraints of a computer.

2. Electronic flow meter for more accurate flow recording.

3. Touch control interface independent of the display, more flexible.

4. Unique furnace sealing provides excellent experimental conditions.

5. Automatic switching of two atmosphere gas lines, fast switching speed and short stabilization time; an additional protective gas input is provided.

6. Simple and easy-to-operate software.

7. Standard substances (indium, tin, zinc) are provided for user self-calibration of temperature.

The video is currently in production, please stay tuned......

Shandong Horde Electronic Technology Co., Ltd.

  • Phone:008619053605658

  • Email:salesmanager@hordeinstrument.com

whatsapp

Copyright © Shandong Horde Electronic Technology Co., Ltd. All Rights Reserved.
Disclaimer: This website cannot identify the intellectual property rights of uploaded images or text. If there is any infringement, please notify us in a timely manner, and this website will review and delete it as soon as possible.

Phone

008619053605658

Contact
Top