Portable Ferrous Content Meter measures the ferromagnetic particle concentration in oil within 10 seconds and outputs ppm data. Widely used for machinery wear monitoring and industrial equipment condition assessment....
The concentration of ferromagnetic particles in oil is one of the most intuitive indicators of equipment wear. The HD-TL200 portable ferrous particle analyzer can detect the concentration of ferrous particles in oil within 10 seconds, simply and intuitively outputting concentration data in ppm (mg/kg). With features such as compact portability, fast detection, small sample consumption, simple operation and high data repeatability, it helps users quickly identify abnormal wear of equipment whether in an industrial site or a laboratory.
The core of the HD-TL200 consists of two sets of precisely wound electromagnetic induction coils. One coil is used for real-time detection, and the other serves as a reference comparison. When a 5 mL sample test tube is inserted into the instrument, ferromagnetic particles such as iron, nickel and cobalt cause a change in the magnetic field of the induction coil. This change is linearly related to the concentration of ferromagnetic particles in the lubricating oil, thereby realizing the detection of ferromagnetic particle concentration in oil from nanometer to millimeter scale.
Oil laboratories, ships, power generation, mining, metallurgy, papermaking, cement, chemical and other engineering machinery and equipment, and transport vehicles.
Gear oil, transmission oil, engine oil, compressor oil, hydraulic oil, etc.
1. Complies with ASTM D8120.
2. Fast detection in 10 seconds.
3. Wide detection range of 0-2000 ppm (mg/kg), expandable to 0-20000 ppm.
4. Unaffected by oil aging, moisture or air bubbles.
5. Requires only a 5 mL sample.
6. Measurement repeatability of <5 ppm.
7. Detects ferromagnetic particles from nanometer to millimeter scale.
8. Compact and portable, with no chemical consumables.
| Detection content | Ferromagnetic particle concentration in lubricating oils and greases in ppm (μg/kg) |
| Display | 4.3-inch touchscreen |
| Detection range | HD-TL200: 0~2000 ppm; HD-TL200+: 0~20% (0~200000 ppm) |
| Data storage | 2000 test records |
| Sensitivity | 1 ppm @ 0~2000 ppm; 5 ppm @ 2000~200000 ppm; 0~500 ppm: <2% or 5 ppm, whichever is greater |
| Applicable fluids | Grease, gear oil, transmission oil, engine oil, compressor oil, hydraulic oil, etc. |
| Repeatability | 500~2000 ppm: <1% or 10 ppm, whichever is greater; 2000~200000 ppm: <0.8% |
| Detection time | <10 s |
| Sample volume | 5 mL |
| Particle response | Nanometer to millimeter scale (ESD equivalent diameter) |
| Calibration accessory | 3 calibration samples included |
| Other accessories | 50 × 5 mL test tubes |
| Working power supply | DC 24 V, 0.25 A |
| Working temperature | 10~45 °C |
| Working humidity | 10~90% RH, non-condensing |
| Storage temperature | -20~70 °C |
| Dimensions | 20 cm × 19 cm × 14 cm |
| Weight | <2.0 kg |
| Electromagnetic compatibility | EN 61000, EN 61326 |
| Compliant standard | ASTM D8120 |