Airborne Particle Counter (Laser Airborne Particle Counter) measures dust particles in six size channels and judges the cleanliness level per GB/T6167-2007. Widely used for workshop cleanliness testing, cleanrooms, hospitals and electronics industries....
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
The intelligent airborne particle counter is an instrument whose confidence-level measurement and verification standard follows the conditions specified in the national standard GB/T 6167-2007 and the calibration specification JJF 1190-2008, and which automatically determines the cleanliness level. One sampling can simultaneously measure the number of dust particles in six particle size channels and simultaneously display the number and variation of particles in the six channels.
It satisfies the cleanliness testing needs of the pharmaceutical industry (pharmaceutical factories, drug inspection institutes, hospitals, etc.); the electronics industry (semiconductor factories, precision machinery production and processing, etc.); the food hygiene industry (fine processing of dairy products, plastic-sealed meat products, condiments, agricultural products, etc.); manufacturers of filters (inspecting filter quality and efficiency, etc.); and cleanrooms (zones) required for precision processing and precision testing in optics, aerospace and other fields.
| Parameter | Specification |
|---|---|
| Sampling volume | 2.83 L/min (0.1 cfm) |
| Six particle size channels | 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 (μm) |
| Maximum sampling concentration | 35000 particles/L (particle size ≥ 0.5 μm) |
| Self-cleaning time | ≤ 10 min |
| Particle size distribution error | 0.5 μm, 5 μm: ≤ ± 30% |
| Particle concentration indication error | 0.5 μm: ≤ ± 30% |
| Measurement repeatability relative error | < 10% |
| Flow error | ≤ ± 5% |
| Sampling period | Arbitrarily settable from 1 s to 30,000 s |
| Operating environment | Working environment 15-35℃, 10-85%RH |
| Alarm function | Particle count alarm realized by setting alarm parameters |
| Sampling delay | ≤ 60 s |
| Measurement interval | ≤ 60 s |
| Sampling mode | Single, ≤ 99 times (user-defined), continuous |
| Light source | Imported semiconductor laser diode (service life more than 30,000 hours) |
| Display | 7-inch color touch screen |
| Data memory capacity | 10,000-100,000 records, queryable and exportable |
| Design working time | 8 hours |
| Charging time | 2 hours |
| Average power | 7.5 W |
| Built-in battery | 74 Wh (16.8V / 5 Ah) |
| Charger | 50 W (16.8V / 3 A) |
| Dimensions | 185 × 220 × 200 mm (D × W × H) |
| Bare weight | 3.25 kg |
| Instrument packaging dimensions | 460 × 265 × 420 mm |
| Packaging weight | 8.6 kg |