Airborne Particle Counter (Laser Airborne Particle Detector) measures dust particles in six size channels and judges the cleanliness level per GB/T6167-2007. Widely used for cleanrooms, pharmaceutical plants, hospitals and electronics industries....

  • Image and Text Display
  • Technical Specifications
  • Product video

Product Introduction

The intelligent airborne particle counter is an instrument whose confidence-level measurement and verification standard follows the conditions specified in the national standard GB/T 6167-2007 and the calibration specification JJF 1190-2008, and which automatically determines the cleanliness level. One sampling can simultaneously measure the number of dust particles in six particle size channels and simultaneously display the number and variation of particles in the six channels.

Application Range

It satisfies the cleanliness testing needs of the pharmaceutical industry (pharmaceutical factories, drug inspection institutes, hospitals, etc.); the electronics industry (semiconductor factories, precision machinery production and processing, etc.); the food hygiene industry (fine processing of dairy products, plastic-sealed meat products, condiments, agricultural products, etc.); manufacturers of filters (inspecting filter quality and efficiency, etc.); and cleanrooms (zones) required for precision processing and precision testing in optics, aerospace and other fields.

Basis Standards

  • National standard: GB/T 6167-2007 Test method for performance of airborne particle counters
  • Cleanliness grade determination standards: ISO 14644, FS 209E, new GMP (dynamic/static)
  • Calibration specification: JJF 1190-2008

Airborne Particle Counter Features

  1. Android 7.1 system for a better interactive operating experience.
  2. Built-in temperature and humidity sensor and ambient air pressure sensor.
  3. The core component, the light-scattering sensor, significantly eliminates stray light in the cavity and greatly improves the signal-to-noise ratio, making measurement more accurate and reliable.
  4. Calculated at 95% confidence level (UCL); counting results can be converted into per m³ and per ft³.
  5. Automatically determines the cleanliness grade according to the cleanliness grade division and regulations, convenient to use.
  6. Multiple collection information configurations can be customized and quickly recalled.
  7. Audit trail function: queryable, traceable and controllable, complying with FDA 21 CFR PART 11 (CL10A / CL20A only).
  8. Supports Wi-Fi transmission for wireless data upload; also supports USB flash drive data copy with driver-free plug and play.
  9. Equipped with a supervision platform; detection results can be transmitted wirelessly directly to the platform for short- and long-term cleanliness analysis to assist management.
  10. Built-in new-generation high-speed thermal printer; the required particle size channels can be selected for flexible printing of results.
  11. AC/DC dual use with built-in rechargeable lithium battery, environmentally friendly and durable, supporting more than 6 hours of continuous sampling.
  12. Convenient and simple retractable pan-tilt bracket for easier sampling installation and operation.

Technical Indicators

ParameterSpecification
Sampling volume28.3 L/min (1 cfm)
Six particle size channels0.3, 0.5, 1.0, 3.0, 5.0, 10.0 (μm)
Maximum sampling concentration35000 particles/L (particle size ≥ 0.5 μm)
Self-cleaning time≤ 10 min
Particle size distribution error0.5 μm, 5 μm: ≤ ± 30%
Particle concentration indication error0.5 μm: ≤ ± 30%
Measurement repeatability relative error< 10%
Flow error≤ ± 5%
Sampling periodArbitrarily settable from 1 s to 30,000 s
Operating environmentWorking environment 15-35℃, 10-85%RH
Alarm functionParticle count alarm realized by setting alarm parameters
Sampling delay≤ 60 s
Measurement interval≤ 60 s
Sampling modeSingle, ≤ 99 times (user-defined), continuous
Light sourceImported semiconductor laser diode (service life more than 30,000 hours)
Display7-inch color touch screen
Data memory capacity10,000-100,000 records, queryable and exportable
Design working time8 hours
Charging time3 hours
Design working time8 hours (5 hours of continuous sampling)
Working power25 W
Built-in battery220 Wh (16.8V / 15 Ah)
Charger84 W (16.8V / 5 A)
Dimensions185 × 220 × 200 mm (D × W × H)
Bare weight4.05 kg
Instrument packaging dimensions460 × 265 × 420 mm
Packaging weight9.9 kg
The video is currently in production, please stay tuned......

Shandong Horde Electronic Technology Co., Ltd.

  • Phone:008619053605658

  • Email:salesmanager@hordeinstrument.com

whatsapp

Copyright © Shandong Horde Electronic Technology Co., Ltd. All Rights Reserved.
Disclaimer: This website cannot identify the intellectual property rights of uploaded images or text. If there is any infringement, please notify us in a timely manner, and this website will review and delete it as soon as possible.

Phone

008619053605658

Contact
Top