Products

Model:
HD-CS3
Introduction:
High-frequency Infrared Carbon Sulfur Analyzer measures carbon and sulfur in metals and materials by infrared absorption after high-frequency induction combustion. Widely used in metallurgy, machinery and material analysis....
The high-frequency infrared carbon-sulfur analyzer is a modern analytical instrument developed on the principle that CO2 and SO2 selectively absorb infrared radiation. It consists of an automatic constant-pressure oxygen supply, high-frequency induction heating, infrared detection, a microcomputer, a printer and an electronic balance. When the sample burns in the high-frequency furnace, the CO2 and SO2 generated are carried by oxygen into the infrared detector; the signal is pre-amplified and processed by the computer, and the carbon and sulfur percentages are displayed on the colour monitor and printed out.
| 1 | High-power high-frequency combustion furnace | furnace power of 3.5 kW or more, suitable for steel, alloy steel, ores and other samples; the third-generation furnace head integrates heating, dust back-blowing, dust suction and a cleaning brush; the power is adjusted by the software and saved to the analytical channel, avoiding full-load operation; dual air cooling prevents temperature fluctuation of the infrared cell; intelligent two-zone automatic dust removal combining high-pressure back-blowing, brushing, discharge and collection minimizes dust interference; high-frequency radiation shielding reduces exposure of the operator. |
| 2 | High-precision infrared analysis system | infrared thermostat technology improves reproducibility; automatic switching between the high and low measuring cells; intelligent sleep mode; simplified structure with fewer components; a common chopper motor for the carbon and sulfur cells ensures signal consistency; key components are imported. |
| 3 | New-generation intelligent analysis software | oscilloscope function for fast circuit diagnosis; multi-stage automatic leak detection and remote online diagnosis and control; automatic shutdown and alarm on operating error or abnormal condition; organic carbon analysis supported. |
| Analytical materials | Ferrous metals, non-ferrous metals, graphite, rare earth metals, alloy materials, steel, ores, organic powders, soil, titanium dioxide, new energy materials and other non-metallic materials; organic carbon analysis is also supported |
| Analysis range | Low carbon 0.00001–0.5%; high carbon 0.01–99.99%; low sulfur 0.00001–0.5%; high sulfur 0.01–99.99%; measuring cells selected according to requirements |
| Sensitivity | Carbon: 0.00001%; sulfur: 0.00001% |
| Precision | Carbon: 0.001 mg (1 ppm) or RSD ≤ 0.5%, whichever is met; sulfur: 0.001 mg (1 ppm) or RSD ≤ 0.8%, whichever is met |
| Reference standards | Carbon conforms to ISO 9556 and sulfur to ISO 4935; the instrument also meets JJG 395-2016 "Verification regulation of carbon and sulfur analyzers", GB/T 20123-2006 and ISO 15350-2000 |
| Detection mode | Carbon: non-dispersive infrared absorption with a high-sensitivity carbon cell; sulfur: non-dispersive infrared absorption with a high-sensitivity sulfur cell |
| Analysis time | Adjustable within 45 s; the end point is controlled by both absorbance and time |
| High-frequency furnace power | 3.5 kW / 20 MHz, dual air-cooling system, quick-release combustion tube assembly |
| Power adjustment | Power adjustment function, automatically matched to the analytical method |
| Dust removal | Intelligent two-zone fully automatic dust removal integrating high-pressure back-blowing, brush cleaning, dust discharge and dust collection |
| Infrared thermostat | Automatic constant-temperature system, temperature difference ± 0.5 °C |
| Analysis parameters | Flexibly changeable according to the sample; a separate calibration curve can be set for each analytical condition |
| Calibration | Single-point, multi-point and linear calibration; first- to fifth-order calibration curves can be established |
| Heating temperature | 2000 °C, steplessly adjustable |
| Intelligent sleep | The instrument automatically cuts power and gas when idle |
| Data interface | USB, RS232, network port and other options |
| Required gas | Oxygen, 99.5% purity, 35 psi (2.41 bar) ± 10% |
Products
Services
About
Contact
Copyright © Shandong Horde Electronic Technology Co., Ltd. All Rights Reserved.
Disclaimer: This website cannot identify the intellectual property rights of uploaded images or text. If there is any infringement, please notify us in a timely manner, and this website will review and delete it as soon as possible.
008619053605658