Products

Model:
HD-CS4
Introduction:
High-frequency Infrared Carbon Sulfur Analyzer measures carbon and sulfur in metals and materials by infrared absorption after high-frequency induction combustion. Widely used in metallurgy, machinery and material analysis....
The high-frequency infrared carbon-sulfur analyzer is a modern analytical instrument developed on the principle that CO2 and SO2 selectively absorb infrared radiation. It consists of an automatic constant-pressure oxygen supply, high-frequency induction heating, infrared detection, a microcomputer, a printer and an electronic balance. When the sample burns in the high-frequency furnace, the CO2 and SO2 generated are carried by oxygen into the infrared detector; the signal is pre-amplified and processed by the computer, and the carbon and sulfur percentages are displayed on the colour monitor and printed out.
| 1 | Full stainless steel gas path | all plastic tubing has been replaced by stainless steel tubing, making the gas path compact, easy to troubleshoot and more corrosion resistant; the lower specific surface area reduces adsorption and improves precision for low-content samples. |
| 2 | Vacuum ceramic capacitor | higher-frequency power release and better long-term stability. |
| 3 | Preheating function | the high-frequency generator starts before the oxygen is admitted, which solves the sample-blowing problem caused by the large top oxygen demand of low-density, hard-to-burn samples. |
| 4 | Double-layer infrared cell isolation | minimizes the effect of ambient temperature changes on long-term stability. |
| 5 | Third-generation furnace head | integrates internal heating, dust back-blowing, dust suction and a cleaning brush. |
| 6 | Hollow furnace tail | leaves no hiding place for dust, eliminates blind zones during dust suction and conveys dust directly from the furnace head to the dust collection box. |
| 7 | Patented cleaning brush | instead of the conventional edge air intake, oxygen passes through a large central channel. |
| 8 | Automatic quartz combustion tube loading | an internationally leading feature comparable to top European and American products. |
| 9 | Improved natural convection cooling | the electronic tube stays below 80 °C during continuous operation, which slows down ageing and greatly extends the service life. |
| 10 | Automatic sample feeding system (optional) | automatically judges the reliability of the test data. |
| 11 | Whole-machine condition monitoring | current, voltage, gas flow and gas pressure are monitored continuously, with an alarm when the temperature is overloaded. |
| Analytical materials | Ferrous metals, non-ferrous metals, alloy materials, rocks and minerals, soil, titanium dioxide, new energy materials and other non-metallic materials; organic carbon analysis is also supported |
| Analysis range | Low carbon 0.00001–0.5%; high carbon 0.01–99.99%; low sulfur 0.00001–0.5%; high sulfur 0.01–99.99% |
| Sensitivity | Carbon: 0.00001%; sulfur: 0.00001% |
| Precision | Carbon: 0.001 mg (1 ppm) or RSD ≤ 0.5%, whichever is met; sulfur: 0.001 mg (1 ppm) or RSD ≤ 0.8%, whichever is met |
| Reference standards | Carbon conforms to ISO 9556 and sulfur to ISO 4935; the instrument also meets JJG 395-2016 "Verification regulation of carbon and sulfur analyzers", GB/T 20123-2006 and ISO 15350-2000 |
| Detection mode | Carbon: non-dispersive infrared absorption with a high-sensitivity carbon cell; sulfur: non-dispersive infrared absorption with a high-sensitivity sulfur cell |
| Analysis time | Adjustable within 45 s; the end point is controlled by both absorbance and time |
| High-frequency furnace power | 3.5 kW / 20 MHz, dual air-cooling system, quick-release combustion tube assembly |
| Power adjustment | Power adjustment function, automatically matched to the analytical method |
| Dust removal | Intelligent two-zone fully automatic dust removal integrating high-pressure back-blowing, brush cleaning, dust discharge and dust collection |
| Infrared thermostat | Automatic constant-temperature system, temperature difference ± 0.5 °C |
| Analysis parameters | Flexibly changeable according to the sample; a separate calibration curve can be set for each analytical condition |
| Gas line | Full mirror-finish stainless steel lines, highly corrosion resistant, leak-free over long periods |
| Capacitor type | High-precision vacuum ceramic capacitor |
| Equipment monitoring | A 12-inch touch screen displays and controls the equipment in real time and monitors gas pressure, voltage, gas flow and current, with an automatic alarm when a rated value is exceeded |
| Operating system | Dual-microcomputer operating system: a 12-inch touch-screen computer is built into the cabinet, so the instrument can be operated and monitored without an external analysis computer |
| Calibration | Single-point, multi-point and linear calibration; first- to fifth-order calibration curves can be established |
| Heating temperature | 2000 °C, steplessly adjustable |
| Intelligent sleep | The instrument automatically cuts power and gas when idle |
| Data interface | USB, RS232, network port and other options |
| Required gas | Oxygen, 99.5% purity, 35 psi (2.41 bar) ± 10% |