Scanning Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES)

Model:

HD-ICP1

Introduction:

Scanning Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) analyzes major, minor and trace elements with high precision and wide dynamic range. Widely used for environmental, geological, metallurgy and material analysis....

  • Image and Text Display
  • Technical Specifications
  • Product video

I. Application Range and Features

The inductively coupled plasma optical emission spectrometer (ICP-OES) has outstanding anti-interference and analytical detection performance. It is used in high-end analysis needs such as R&D and testing, and can conveniently perform qualitative, semi-quantitative and precise quantitative analysis. It is an ideal instrument for simultaneous analysis of major, minor and trace inorganic elements.

High precision and wide dynamic range: supports precise analysis from major to trace elements with outstanding anti-interference capability, meeting diverse analysis needs.

Flexible RF power supply system: power continuously adjustable from 400 W to 1800 W; low-power mode reduces argon consumption and adapts to the analysis of a variety of complex samples.

Multi-functional observation modes: provides axial, radial and other observation methods; the signal automatic attenuation function simplifies sample handling and improves detection efficiency.

Advanced optical design: a thermostatic optical system and a high-resolution grating ensure long-term stability and accuracy and effectively resist the influence of ambient temperature changes.

Friendly software platform: a full Chinese graphical interface that is easy to operate, supports remote diagnosis and maintenance, and facilitates the generation of test reports and data management.

Optimized sample introduction system: adopts MFC-controlled gas flow and optional diversified sample introduction systems (such as high-salt-resistant or HF-resistant) to adapt to various sample conditions.

II. Working Environment

Voltage220 V AC ±10%
Room temperatureAmbient temperature (25-50) ℃
Relative humidity(50-70)% RH

III. Technical Specifications

Plasma RF sourceSolid-state RF power supply, power range 400 W-1800 W, continuously adjustable, with stronger sample adaptability.
RF generatorSelf-excited RF generator: extremely fast oscillation matching at startup, strong power load capacity, able to directly analyze 100% toluene samples; automatic oscillation tuning; water-cooled heat dissipation; adapts to power switching for complex sample analysis, shock-resistant and interference-resistant and reliable.
Low-power modeProvides 400 W ultra-low power standby with a small-flow mini torch, reducing argon consumption by more than 50%; argon consumption in standby ≤5 L/min.
Plasma observationConfigurable in four observation modes: axial, radial, bidirectional and simultaneous bidirectional. The vertical torch avoids high-salt deposition, extending torch life and reducing torch consumable costs. Axial observation uses a metal cold cone to remove the tail flame, obtaining higher sensitivity with no need for an air compressor, saving external accessories and consumables. The radial observation plasma observation position is adjustable, providing stronger anti-interference capability for the analysis needs of different elements.
Signal automatic attenuationWith axial and radial attenuation functions, able to attenuate samples with a content above 30%, so that high-content elements in a sample can be analyzed in a single run without repeated dilution, reducing sample pretreatment difficulty and supporting simplified analysis.
RF frequency27.12 MHz/40.68 MHz (optional), efficiency greater than 75%.
RF power / frequency stabilityRF power stability ≤0.05%; RF frequency stability ≤0.01%.
Working coilCopper tube inner diameter 25 mm / 3 turns.
Quartz torchThree-concentric type, outer diameter 20 mm.
NebulizerUniversal high-salt-resistant nebulizer, outer diameter 6-6.2 mm.
Spray chamberDouble-barrel / cyclone spray chamber.
Optical systemThermostatic optical system, C-T optical path design with few reflections and low light energy loss. All optical elements are sealed in a thermally balanced light chamber; the main unit and light chamber are thermally isolated, better resisting external ambient temperature changes; the light chamber is precisely thermostatically controlled at 35 ℃ ±0.1 ℃.
Grating spectroscopyStable and efficient fully fixed high-resolution grating spectroscopy system, strong shock resistance, stable and reliable.
Wavelength range165-850 nm, three-grating full-wavelength coverage.
MonochromatorUses a concave mirror instead of a convex lens as the optical focusing element, solving the chromatic aberration problem caused by different focal points for different elements and improving optical system efficiency.
Grating typeIon-etched holographic grating; the more lines, the higher the resolution (2400 L / 3600 L / 4320 L gratings available).
Wavelength calibrationAt each ignition, only Ar spectral lines are used for automatic spectral position calibration, ensuring the correctness of the analysis wavelength, with no wavelength calibration solution required.
Argon line real-time calibrationUses interference-free argon characteristic lines for real-time correction of slight spectral shifts, achieving optimal spectral integration, ensuring long-term stability and effectively eliminating the influence of spectral drift on measurement.
Purged light chamberFor the determination of wavelengths below 190 nm, nitrogen is selected for optical path purging.
Stray light≤2.0 mg/L (10000 mg/L Ca solution measured at As 188.980 nm).
Optical resolution0.007 nm (3600-line grating); 0.014 nm (2400-line grating); 0.004 nm (4320-line grating).
DetectorDedicated imported Hamamatsu detector (R928/R212UH).
Anti-saturation overflowBack-illuminated anti-overflow design for each spectral line, completely eliminating the spectral-line saturation overflow problem, with no need to worry about the influence of spectral-line saturation on adjacent lines.
Intelligent integrationSynchronous signal-background acquisition, with integration time depending on the light intensity of the spectral line, automatically calculating the optimal integration time, obtaining both high-intensity and weak signals at the best signal-to-noise ratio, broadening the dynamic range so that high- and low-content elements can be detected simultaneously, avoiding repeated sample dilution.
Detector surfaceNo light-conversion chemical coating on the detector surface, so detector replacement due to coating aging will not occur.
Sample introduction systemStandard quartz concentric nebulizer, quartz cyclone spray chamber, classic quartz double-barrel spray chamber, integrated quartz torch.
Optional argon online dilutionHigh-precision MFC-controlled argon efficiently dilutes high-salt samples above 10%, allowing direct injection.
Optional sample introductionHigh-salt-resistant or HF-resistant sample introduction system, organic sample introduction system and hydride generator.
Gas path controlIn the sample introduction system, the carrier gas, plasma gas and auxiliary gas are all controlled by advanced mass flow controllers (MFC), with advantages such as continuously adjustable flow and stable flow, ensuring the stability of the sample introduction system and providing a strong guarantee for light source stability.
Analysis speedApprox. 25 elements per minute.
Sample consumptionOnly 5 mL required.
Linear dynamic range≥10⁵ (measured at Mn 257.6 nm, correlation coefficient ≥0.999).
PrecisionMeasuring a 1 ppm or 10 ppm multi-element mixed standard solution, RSD of ten repeated measurements ≤0.5%.
StabilityMeasuring a 1 ppm or 10 ppm multi-element mixed standard solution, 8-hour long-term stability RSD ≤1%.
Detection limits (µg/L, per JJG 768-2005)Zn 213.856 nm: 0.5; Ni 231.604 nm: 1; Mn 257.610 nm: 0.5; Cr 267.716 nm: 1; Cu 324.754 nm: 1; Ba 455.403 nm: 0.1.
Preheating timeFrom standby to plasma ignition less than 3 minutes.

IV. Accessory System

Computer system: 4 GB RAM, 1 TB hard disk, with DVD and keyboard/mouse.

Cooling circulating water system: cooling capacity 1600 W, temperature range 20-50 ℃.

Monochrome laser printer.

Single-phase AC voltage stabilizer (15 KVA, accuracy 1%, input voltage 160 V-280 V, output voltage 220 V).

UPS power supply 10 KVA.

Grounding copper strip.

The video is currently in production, please stay tuned......

Shandong Horde Electronic Technology Co., Ltd.

  • Phone:008619053605658

  • Email:salesmanager@hordeinstrument.com

whatsapp

Copyright © Shandong Horde Electronic Technology Co., Ltd. All Rights Reserved.
Disclaimer: This website cannot identify the intellectual property rights of uploaded images or text. If there is any infringement, please notify us in a timely manner, and this website will review and delete it as soon as possible.

Phone

008619053605658

Contact
Top