Introduction:
Online Trace Dissolved Oxygen Meter measures ppb-level dissolved oxygen in water and gas based on Clark electrode technology. Widely used for power plant and boiler water testing and semiconductor ultrapure water trace oxygen detection....
The measurement is based on Clark electrode technology and is designed specifically for the determination of trace (ppb-level) oxygen in water and gases.
Detection and analysis of low-concentration dissolved oxygen in power plant and boiler feed water, and trace oxygen measurement in semiconductor ultrapure water.
1. High-resolution colour touch screen: 800 × 480 graphic dot matrix with a Chinese menu, convenient and real-time operation.
2. Curve display: measurement results are shown as real-time digits together with a trend curve, intuitive and convenient.
3. Multi-channel configuration: up to 4 dissolved oxygen sensors can be connected, saving cost for the customer.
4. High precision and high sensitivity: detection limit 0.001 μg/L, resolution 0.001 μg/L.
5. Fast response and measurement: generally only 3 minutes from air-saturated oxygen concentration to the μg/L level.
6. Long-life electrode: the long electrode service life reduces the cost of frequent electrode replacement.
7. Long maintenance interval and low-cost consumables: normal operation requires electrode maintenance only every 6–8 months.
8. Large-capacity data storage: 1.5 million data records with historical curve query.
| Parameter | RY500 | RY500M | RY500L |
|---|---|---|---|
| Measuring range | 0.01 μg/L–20 mg/L | 0.1 μg/L–20 mg/L | 1 μg/L–20 mg/L |
| Resolution | 0.01 μg/L | 0.1 μg/L | 1 μg/L |
| Accuracy | ±1 % FS or ±0.5 μg/L (whichever is greater) | ±1.5 % FS or ±1.0 μg/L (whichever is greater) | ±2 % FS or ±2 μg/L (whichever is greater) |
| Response time | T90 < 20 s | T90 < 30 s | T90 < 30 s |
| Display | 800 × 480 colour touch screen, Chinese menu (manual switch to English) | ||
| Channels | Up to 4 dissolved oxygen sensors | ||
| Sample temperature | 0–50 °C; resolution 0.1 °C / 0.01 °C | ||
| Sample flow rate | 100–400 mL/min | ||
| Electrode service life | More than 5 years (depending on the sample) | ||
| Operating environment | Temperature 0–50 °C; relative humidity ≤ 90 % | ||
| Current output | 0/4–20 mA (two isolated outputs) | ||
| Current output accuracy | ±0.01 mA | ||
| Maximum load resistance | 800 Ω (20 mA max.) | ||
| Alarm output | Two alarm outputs | ||
| Power supply | AC 220 V; DC 24 V | ||
| Data storage | 1.5 million data records | ||
| Communication | RS485 | ||
| Weight | 800 g | ||
| Installation | Panel cut-out / wall mounting / pipe mounting | ||
| Cut-out size | 138 × 138 mm | ||
Model HD-RY500L corresponds to the RY500L column above; the other columns are listed for model comparison.