Introduction:
Automatic Distillation Apparatus (Intelligent Integrated Distillation Apparatus) performs automatic distillation for laboratory sample pretreatment. Widely used for water quality, soil, food and environmental sample distillation with high automation and efficiency....
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
.jpg)
The fully automatic integrated distillation instrument is an intelligent, efficient high-tech product designed for modern laboratories, aiming to significantly improve experimental efficiency and data accuracy through highly automated and intelligent operation. The instrument integrates advanced cooling water circulation technology, independent multi-channel control, a 7-inch LCD touchscreen, automatic cleaning, automatic leak-tightness check, back-blowing and remote OTA system upgrade, providing users with a one-stop distillation solution.
1. High degree of automation: built-in cooling water circulator realizes fully automatic operation, improving experimental efficiency and accuracy.
2. Independent multi-channel control: 3 heating units are independent of each other, each equipped with anti-siphon, excessive-distillation prevention and anti-dry-boiling protection.
3. Intelligent touch control: 7-inch LCD touchscreen supports user-defined storage of up to 1000 distillation methods, with 18 built-in national standard methods.
4. Precise control: high-efficiency ceramic far-infrared heating bowl with dual time and weight end-point control, ensuring distillation accuracy of 0.1 mL.
5. Safe and reliable: integrated automatic cleaning, automatic leak-tightness check and back-blowing functions ensure long-term stable operation.
6. Easy operation: one-key start, supports remote OTA system upgrade to meet the needs of different users.
| Parameter | ZLD | ZLE |
|---|---|---|
| Operation | 7-inch LCD color capacitive touchscreen | |
| Heating method | Far-infrared ceramic heating | |
| Heating units | 3 channels | |
| Heating flask | 500 mL spherical ground-neck round-bottom flask | |
| Control mode | Single-position single control, one-key multi-position control | |
| Functions | Anti-siphon, excessive-distillation protection, automatic anti-dry-boiling, closed-loop flow rate control, closed-loop temperature control, automatic cleaning, automatic leak-tightness check, multiple distillation methods, built-in cooling water chiller | |
| Distillation accuracy | 0.1 mL | |
| Distillation temperature | Room temperature ~ 400 °C | |
| Distillation speed | 12 mL/min (varies with output power) | |
| Heating-up time | Time from power-on to boiling: ≤6 min (at 25 °C room temperature, 600 W per position, 250 mL tap water in a 500 mL flask) | |
| Time control | 0-999 min | |
| Maximum heating power | 600 W per position | |
| End-point control | Time | Distilled volume + time |
| Per-position control | Time + temperature | Distilled volume + time + density + weighing calibration parameters + temperature |
| Method save/recall | Up to 1000 distillation methods storable, with 18 built-in common national standard methods | |
| Back-blowing | Automatic back-blowing to empty circulating water in the condenser | |
| One-key automatic cleaning | Yes (built-in automatic cleaning pump, no need to disassemble the flask; distillation pipeline can be cleaned with one key after the experiment) | |
| Automatic anti-dry-boiling | Temperature method + flow rate method + volume method | |
| Closed-loop flow control | Yes | |
| Remote system upgrade | Yes | |
| Automatic leak-tightness check | Yes | |
| Leak-tightness test pressure | ≥-80 kPa | |
| Cooling method | Built-in cooling water chiller, closed-loop temperature control, controllable 5 °C ~ 25 °C | |
| Chiller cooling capacity | 1.5 kW | |
| Distillate receiving flask | 250 mL × 3 (standard) | |
| Rated power | 2000 W | |
| Power supply | 220 V 50 Hz | |
| Leakage protection | Yes | |
| Dimensions (mm) | 560×533×634 | |
| Weight (kg) | 46 (main unit + glassware) | |
GB/T 5750.5-2023 Standard examination methods for drinking water - Non-metallic inorganic indicators
HJ 484-2009 Water quality - Determination of cyanide - Volumetric and spectrophotometric methods
HJ 503-2009 Water quality - Determination of volatile phenols - 4-AAP spectrophotometric method
HJ 502-2009 Water quality - Determination of volatile phenols - Bromination volumetric method
HJ 1226-2021 Water quality - Determination of sulfide - Methylene blue spectrophotometric method
HJ 1191-2021 Water quality - Determination of azide - Spectrophotometric method
HJ 535-2009 Water quality - Determination of ammonia nitrogen - Nessler reagent spectrophotometric method
HJ 536-2009 Water quality - Determination of ammonia nitrogen - Salicylic acid spectrophotometric method
HJ 537-2009 Water quality - Determination of ammonia nitrogen - Distillation-neutralization titration method
HJ 745-2015 Soil - Determination of cyanide and total cyanide - Spectrophotometric method
HJ 998-2018 Soil and sediment - Determination of volatile phenols - 4-AAP spectrophotometric method
HJ 833-2017 Soil and sediment - Determination of sulfide - Methylene blue spectrophotometric method
HJ 717-2014 Soil quality - Determination of total nitrogen - Kjeldahl method
GB 8538-2022 National food safety standard - Examination methods for drinking natural mineral water
GB 5009.36-2023 National food safety standard - Determination of cyanide in foods
Products
Services
About
Contact
Copyright © Shandong Horde Electronic Technology Co., Ltd. All Rights Reserved.
Disclaimer: This website cannot identify the intellectual property rights of uploaded images or text. If there is any infringement, please notify us in a timely manner, and this website will review and delete it as soon as possible.
008619053605658