Ambient Air Comprehensive Sampler collects particulate matter (TSP, PM10, PM2.5) and gas components (SO2, NOx) in the atmosphere. Widely used for environmental monitoring stations, air quality assessment and atmospheric research....

  • Image and Text Display
  • Technical Specifications
  • Product video

Product Overview

The HD-H7 (Model H7) Environmental Air Integrated Sampler is an indispensable instrument for capturing particulate matter (TSP, PM10, PM2.5, etc.) and various gas components (SO2, NOx, etc.) from the atmosphere. During development, opinions from experts and a wide range of users were extensively collected, and high technologies in the fields of computers, sensors and new materials were applied, resulting in reliable quality, stable performance and a long service life.

Application Range

The sampler uses the solution absorption method to collect various harmful gases in ambient and indoor air, and uses the filter-membrane gravimetric method to capture total suspended particulates (TSP) and inhalable particulates (PM10 or PM2.5), fluoride, heavy metals, SVOC, etc. It can be used by environmental protection, health, labor, safety supervision, military, scientific research and education departments for routine and emergency monitoring of gaseous substances and aerosols.

Executive Standards

  • HJ 618-2011 Ambient air - Determination of PM10 and PM2.5 - Gravimetric method
  • HJ 656-2013 Technical specification for manual monitoring of ambient air PM2.5 (gravimetric method)
  • HJ 93-2026 Technical requirements and test methods for ambient air particulate matter (PM10 and PM2.5) samplers
  • GB/T 18883-2022 Indoor air quality standard
  • HJ/T 374-2007 Technical requirements and test methods for total suspended particulate samplers
  • HJ/T 375-2007 Technical requirements and test methods for ambient air samplers
  • HJ/T 376-2007 Technical requirements and test methods for 24-hour thermostatic automatic continuous ambient air samplers
  • JJG 956-2013 Air samplers
  • JJG 943-2011 Total suspended particulate samplers
  • HJ 955-2018 Ambient air - Determination of fluoride - Filter membrane sampling/fluoride ion selective electrode method
  • HJ 657-2013 Ambient air and waste gas - Determination of lead and other metal elements in particulate matter - Inductively coupled plasma mass spectrometry
  • HJ 644-2013 Ambient air - Determination of volatile organic compounds - Sorbent tube sampling-thermal desorption/gas chromatography mass spectrometry
  • HJ 583-2010 Ambient air - Determination of benzene homologues - Solid adsorption/thermal desorption-gas chromatography

Technical Features

  1. Multiple uses in one instrument: free combination of medium-flow sampling, low-flow heating, low-flow thermostatic and micro-flow sampling modes, with independent control and flexible configuration.
  2. 7.0-inch wide-temperature high-brightness color touch display, clearly visible in sunlight, with a beautiful interface and convenient operation.
  3. Dual operation mode of keys and touch, adapting to the operating habits of various users.
  4. Optional built-in rechargeable lithium battery; a single charge provides more than 4 hours of working time.
  5. Massive data storage: more than 100,000 data records.
  6. Particulate sampling flow (10-130) L/min, settable; with the corresponding sampling head, it can be used for sampling fluoride, heavy metals and SVOC.
  7. Imported sampling pump with high load capacity, low noise, stable flow and long service life.
  8. Automatically measures pre-meter temperature and pre-meter pressure and automatically calculates the standard-condition sampling volume.
  9. Small size, light weight and convenient to carry.
  10. Automatic data protection when power fails during sampling; sampling continues after power is restored.

Technical Indicators

Main parameterRangeResolutionAccuracy
H6 (high negative pressure type) - particulate sampling flow(10-130) L/min0.1 L/minBetter than ± 2.0%
H7 (high negative pressure + dual-channel constant-flow type) - particulate sampling flow(10-130) L/min0.01 L/minBetter than ± 2.0%
Channel A, Channel B(0.100-1.000) L/min0.001 L/minBetter than ± 2.5%
Flow stabilityBetter than ± 2.0%
Flow repeatabilityBetter than ± 2.0%
Load capacityOvercomes -20 kPa resistance at 100 L/min flow
Pre-meter pressure(-35 to 35) kPa0.01 kPaBetter than ± 2.5%
Atmospheric pressure(50-110) kPa0.001 kPaBetter than ± 500 Pa
Sampling time1 min to 99 h 59 min1 minBetter than ± 0.2%
Equal-interval sampling timeArbitrarily set within 99 h 59 min
Equal-interval sampling times1-99 times
Noise< 60 dB(A)
Dimensions (W × D × H)165 × 330 × 275 mm
WeightAbout 6.5 kg
Power supplyAC 220V ± 10% 50Hz
Power consumption< 100 W
Sampling head - PM2.5 cutting characteristicsDa50 = (2.5 ± 0.2) μm and g = (1.2 ± 0.1) μm
Sampling head - PM10 cutting characteristicsDa50 = (10 ± 0.5) μm and g = (1.5 ± 0.1) μm
Inlet velocity0.3 m/s
b/a0.625
Effective filter diameterφ80 mm
ConnectorM20 × 1.5
The video is currently in production, please stay tuned......

Shandong Horde Electronic Technology Co., Ltd.

  • Phone:008619053605658

  • Email:salesmanager@hordeinstrument.com

whatsapp

Copyright © Shandong Horde Electronic Technology Co., Ltd. All Rights Reserved.
Disclaimer: This website cannot identify the intellectual property rights of uploaded images or text. If there is any infringement, please notify us in a timely manner, and this website will review and delete it as soon as possible.

Phone

008619053605658

Contact
Top