Flame Photometer, based on atomic emission (flame photometry), measures the concentration of alkali and alkaline-earth metals such as potassium, sodium, calcium and lithium in solution. Widely used in agriculture, soil testing, cement, ceramics and glass industries....

  • Image and Text Display
  • Technical Specifications
  • Product video

I. Product Overview

The flame photometer is a relative measuring instrument based on the principle of atomic emission spectrometry (flame color reaction), mainly used to determine the concentration of alkali metal and alkaline earth metal elements (such as potassium, sodium, calcium, lithium, etc.) in solutions.

The concentration value of the tested sample is the relative value with respect to the concentration of a standard solution under the same test conditions. Before testing, a set of standard solutions must first be prepared, then a calibration operation is carried out to draw a standard curve manually or via the instrument, and finally the tested sample is measured to obtain its concentration value or other required calculation data.

II. Applications

It can be widely used in environmental monitoring, agriculture and fertilizer industry, food industry, chemical manufacturing, medical and clinical diagnosis, geology, mining and new energy industries, and other fields with sample pretreatment requirements.

III. Executive Standards (Partial)

JJG630-2007 Flame Photometers

LY/T 1246-1999 Determination of Exchangeable Potassium and Sodium in Forest Soil

LY/T 1248-1999 Determination of Exchangeable Sodium in Alkalinized Soil

LY/T 1254-1999 Determination of Total Potassium and Total Sodium in Forest Soil

NY/T 889-2004 Determination of Available Potassium and Slowly Available Potassium in Soil

NY/T 2420-2013 Determination of Total Potassium in Plants by Flame Photometry

IV. Product Features

1. Equipped with a 7-inch color capacitive touch LCD screen, offering convenient operation, a compact size, and stable and reliable performance;

2. Direct concentration reading for five elements (potassium, sodium, lithium, calcium, and barium), with the ability to simultaneously measure any combination of 3 elements;

3. High-transmittance, ultra-low stray-light interference filters reduce background interference to achieve a higher signal-to-noise ratio;

4. One-key ignition function for quick and easy operation, and a built-in silent air pump that eliminates the need for an additional air compressor;

5. Independently developed high-performance atomizer ensures uniform and efficient atomization and smooth, stable sample aspiration, guaranteeing the accuracy and stability of analysis results;

6. Can store standard curve parameters of up to 10 points per set, with a total capacity of 20 sets of curve data;

7. Equipped with a moisture separator to avoid the influence of environmental changes and ensure stable analysis results;

8. Equipped with a flameout protection device, featuring an automatic gas solenoid valve shut-off function after flameout;

9. Equipped with a single-point calibration function to allow quick correction when baseline drift occurs;

10. Equipped with a history record printing function and a USB flash drive export function.

V. Technical Parameters

ModelHD-FP2HD-FP3AHD-FP3BHD-FP3CHD-FP4HD-FP5
Test ElementsNa, KNa, K, CaNa, K, LiNa, K, BaNa, K, Li, CaNa, K, Li, Ca, Ba
Number of Channels233345
Operation ModeTouch screen
Splitting MethodInterference filter
Photoelectric Conversion ElementSilicon photocell
Printing FunctionBuilt-in thermal printer
USB Data OutputHistorical records can be exported via USB flash drive
Concentration Calculation MethodSegmented method, linear fitting, curve fitting
StabilityContinuous injection of standard solution: within 15 s the relative maximum change of the indicated value is ≤ 3%; measured once per minute for 6 times in total, the maximum change of the indicated value is ≤ 8%
Repeatability≤ 2%
Response Time≤ 8 s
Sample Aspiration Rate≤ 6 mL/min
Linearity Error (mmol/L) - SodiumNa ≤ 0.02
Linearity Error (mmol/L) - PotassiumK ≤ 0.003
Linearity Error (mmol/L) - Lithium——Li ≤ 0.007—Li ≤ 0.007—
Linearity Error (mmol/L) - Calcium—Ca ≤ 0.062——Ca ≤ 0.062—
Linearity Error (mmol/L) - Barium———Ba ≤ 0.066—Ba ≤ 0.066
Detection Limit (mmol/L) - SodiumNa ≤ 0.007
Detection Limit (mmol/L) - PotassiumK ≤ 0.0005
Detection Limit (mmol/L) - Lithium——Li ≤ 0.012—Li ≤ 0.012—
Detection Limit (mmol/L) - Calcium—Ca ≤ 0.037——Ca ≤ 0.037—
Detection Limit (mmol/L) - Barium———Ba ≤ 0.066—Ba ≤ 0.066
Detection Range (ppm)Na: 0.18-100; K: 0.15-100; Li: 0.1-30; Ca: 0.1-1000; Ba: 0.1-300
Filter Center Wavelength ErrorNa ≤ 5 nm; K ≤ 7 nm; Li ≤ 7 nm; Ca ≤ 7 nm; Ba ≤ 7 nm
Filter Bandwidth≤ 15 nm
Working ConditionsAmbient temperature: 10 ℃ ~ 35 ℃; relative humidity: ≤ 85%
Main Unit DimensionsLength 460 mm × Width 325 mm × Height 450 mm
Net Weight13 kg ± 1 kg
Power Supply & Power220 V ± 22 V, 50 Hz ± 1 Hz, ≤ 50 W
The video is currently in production, please stay tuned......

Shandong Horde Electronic Technology Co., Ltd.

  • Phone:008619053605658

  • Email:salesmanager@hordeinstrument.com

whatsapp

Copyright © Shandong Horde Electronic Technology Co., Ltd. All Rights Reserved.
Disclaimer: This website cannot identify the intellectual property rights of uploaded images or text. If there is any infringement, please notify us in a timely manner, and this website will review and delete it as soon as possible.

Phone

008619053605658

Contact
Top