Products

Model:
HD-DJ6
Introduction:
Online ORP Sensor measures the oxidation-reduction potential of water with a platinum indicator electrode and RS485 output. Widely used for environmental water testing and industrial water quality monitoring....
The HD-DJ6 online ORP sensor is intended for environmental water quality monitoring and for water samples of high or low ionic strength. It satisfies the requirements of most industrial and environmental applications for continuous online oxidation-reduction potential measurement.
1. The indicating electrode is a platinum sheet, giving high stability.
2. Internal signal isolation technology for strong interference immunity.
3. Good reproducibility, thermal stability and low drift, with a compact body and fast response.
4. 3/4 NPT pipe thread for simple immersion mounting or installation in pipelines and tanks.
5. Stable performance and long service life, with an IP68 waterproof rating.
6. Single-point calibration and RS485 (MODBUS-RTU) output.
| Measurement principle | Electrochemical (salt bridge) |
| Signal output | RS485 (MODBUS-RTU) |
| Measurement range | -1500 mV to +1500 mV |
| Resolution | 1 mV |
| Accuracy | ±6 mV |
| Housing material | ABS |
| Cable length | 5 m standard |
| Power supply | 12–24 VDC (0.4 W @ 12 V) |
| Calibration | Single-point calibration |
| Operating environment | 0–50 °C (non-freezing), < 0.2 MPa |
| Waterproof rating | IP68 |
| Mounting | Immersion mounting, NPT 3/4 thread |
Do not install the sensor upside down or horizontally; install it at an angle of at least 15° to the horizontal.
The cable is a 5-core shielded cable: brown - power supply (12–24 VDC), black - ground (GND), green - RS485-B, yellow - RS485-A. Check the wiring order carefully before powering on, to avoid unnecessary losses caused by incorrect connection. Cable colours may differ slightly between production batches; the colour print on the cable always prevails.
Before first use, or when reusing a sensor that has been stored for a long time, soak the sensitive bulb and the reference junction in 3.3 mol/L KCl solution for more than 2 hours. Rinse with deionised (or distilled) water and dry before use so that impurities are not carried into the sample, and immerse at least one third of the sensor in the liquid under test. Keep the terminal block clean and dry; when the sensor is not in use, clean it and fit the protective cap filled with 3.3 mol/L KCl solution, or store the sensor in a container of the same solution. If the terminal block is soiled, wipe it with anhydrous alcohol and let it dry. Avoid prolonged immersion in deionised water, protein solutions, strong acids, strong alkalis or acidic fluoride solutions, and keep the sensor away from silicone grease. If the platinum sheet is contaminated with grease, wipe it gently with degreased cotton moistened with acetone or alcohol; insoluble inorganic contamination can be removed with 30–50% hydrochloric acid. The platinum sheet may also be wiped once a day with clean soft tissue. When measurement error appears, recalibrate the sensor with a host instrument; if maintenance no longer restores calibration and measurement, replace the sensor.